2D phase encoding imaging pulse sequence with optional diffusion weighting during the echo time. This sequence must be called from the imaging.m context, which would provide H,R,K,G, and F.

Syntax

    mri=phase_enc_2d(spin_system,parameters,H,R,K,G,F)

Parameters

 parameters.t_echo        -  echo time, seconds

 parameters.diff_g_amp    -  [optional] a vector of diffusion gra-
                             dient pair amplitudes in X,Y (T/m) to
                             be active during the echo time

 parameters.pe_grad_amp   -  phase encoding gradient amplitude, T/m

 parameters.ro_grad_amp   -  readout gradient amplitude, T/m

 parameters.pe_grad_dur   -  the duration of the phase encoding 
                             gradient, seconds 

 parameters.ro_grad_dur   -  the duration of the readout gradient
                             seconds

 parameters.image_size    -  number of points in each dimension of
                             the resulting image

Outputs

 mri - MRI image with square sinebell apodisation

See also

epi_2d.m, imaging.m, spiral.m

Magnetic resonance imaging