Difference between revisions of "Dpfgse suppress.m"
(Update function See also links and function index membership) |
(sync with Spinach main f053e432: soft pulse rank parameter is parameters.max_rank) |
||
| Line 21: | Line 21: | ||
parameters.rf_dur_list function | parameters.rf_dur_list function | ||
parameters.rf_phi | parameters.rf_phi | ||
| − | parameters. | + | parameters.max_rank |
parameters.sweep - detection sweep width, Hz | parameters.sweep - detection sweep width, Hz | ||
Latest revision as of 06:40, 30 August 2026
1D pulse-acquire NMR sequence with DPFGSE signal suppression, based on Equation 3 from Stott et al. (https://doi.org/10.1006/jmre.1997.1110); the spatial coordinate is modelled explicitly, and the function should be called through the imaging.m wrapper.
Syntax
fid=dpfgse_suppress(spin_system,parameters,H,R,K,G,F)
Parameters
parameters.rho0 - initial condition parameters.coil - detection state parameters.g_amp - amplitudes of the two gradients, T/m parameters.g_dur - gradient duration, seconds parameters.rf_frq_list - soft pulse parameters that will parameters.rf_amp_list be passed to shaped_pulse_af.m parameters.rf_dur_list function parameters.rf_phi parameters.max_rank parameters.sweep - detection sweep width, Hz parameters.npoints - number of points in the fid parameters.offset - transmitter and receiver offset, Hz
Outputs
fid - free induction decay of what is effectively a
1D pulse-acquire NMR experiment
Notes
At least a hundred points are required in the spatial dimension; increase until the answer stops changing.
See also
imaging.m, dpfgse_select.m, spencosy.m, spendosycosy.m, shaped_pulse_af.m, Built-in_experiments
Version 2.1, authors: Ilya Kuprov, Pavan Lally, Ahmed Allami