Dnp field scan.m
Magnetic field scan steady-state DNP experiment.
Description
Returns the steady-state population of the user-specified state as a function of magnetic field.
Syntax
answer=dnp_field_scan(spin_system,parameters,H,R,K)
This function should normally be called via powder.m or crystal.m wrapper.
Arguments
parameters.mw_pwr - microwave power, Hz
parameters.mw_frq - microwave frequency, Hz
parameters.fields - a vector of B0 field values, T
parameters.rho_eq - thermal equilibrium state
parameters.coil - detection state vector or a horizon-
tal stack thereof
parameters.mw_oper - microwave irradiation operator
parameters.ez_oper - Lz operator on the electrons
parameters.method - 'backslash' to use Matlab's
linear equation solver, 'gmres'
to use ILU preconditioned GMRES
H - Hamiltonian commutation superoperator
R - unthermalised relaxation superoperator
K - chemical kinetics superoperator
Returns
answer - the projection of the equilibrium density matrix
on the user-specified state as a function of the
applied magnetic field
Examples
Cross effect in a system with two spin-1/2 electrons (cross_effect_field_scan_1.m example file in examples/dnp_solids directory):
Solid effect in a system with a spin-7/2 gadolinium ion (solid_effect_field_scan_1.m example file in examples/dnp_solids directory):
Notes
- The relaxation superoperator supplied to this function should NOT be thermalized.
- Thermal equilibrium state is assumed to be the same at all fields in the sweep. Do not use this function for broad magnetic field sweep experiments.
- Backslash method is recommended for small spin systems and GMRES method for large ones.
- This function can be made to run on GPUs. If you would like that functionality added, send an email to Ilya Kuprov.
See also
dnp_freq_scan.m,masdnp.m,solid_effect.m
Revision 3284, authors: Ilya Kuprov